Ceramic capacitive measuring system and built-in breather tube for direct air pressure compensation, the transmitter is designed for use in water applications where improved accuracy for both high and low measuring ranges are requested. Approved for use in explosion hazardous areas - UL Class 1, Division 1, Group A-D.
Ceramic capacitive measuring system and built-in breather tube for direct air pressure compensation, the transmitter is designed for use in water applications where improved accuracy for both high and low measuring ranges are requested. Approved for use in explosion hazardous areas - UL Class 1, Division 1, Group A-D.
Useful for level measurements in ground water bore holes and in-fill sites.
Features
Applications
Nominal Measuring Range | 0 to 3' |
Measuring Principle | Ceramic capacitive, relative pressure |
Temperature Range | -5 to 105°F |
Temp. Deviation, Zero Point | Better than ± 0.01 %/°F |
Temperature Deviation, FS | Better than ± 0.005 %/°F |
Linearity/Stability | Better than ± 0.2 % FS/± 0.2 % FS |
Measurement Accuracy | Better than ± 0.1 % FS @ 50 to 85°F Better than ± 0.2 % FS @ full temperature range |
Long Time Stability | Better than ± 0.1 % FS per year |
Materials | Housing: stainless steel (AISI 316 L) Diaphragm: 96,6 % Al2O3/Cell Packing: Viton |
Supply Voltage | 10 to 30 V DC |
Output Signal | 2-wire 4 - 20 mA (passive transmitter) |
Cable | 2 × AWG 20 (pressure) + 5 x AWG 25 (data), shielded, PUR insulation |
Cable Length | 39' |
Enclosure | NEMA 6P, withstands a static pressure equal to max. overpressure. |
Safety Approvals | UL Class I, II and II, Division 1 Hazardous (Classified) Locations, UL 913 ATEX: II 2G EEx ia IIC T4...T6 |
Test Approvals | CE directive: 89/336/EEC, 2004/108/EC, 1999/EC EN 61326-1, 61326-2-3, EN 55011, 61000-4-2, 61000-4-3, 61000-4-4, 61000-4-5,61000-4-6, 61000-4-8 EN 61000-6-4 2007-02-19, EN 61000-6-2 2005-09-08 |
Nominal Measuring Range | 0 to 3' |
Measuring Principle | Ceramic capacitive, relative pressure |
Temperature Range | -5 to 105°F |
Temp. Deviation, Zero Point | Better than ± 0.01 %/°F |
Temperature Deviation, FS | Better than ± 0.005 %/°F |
Linearity/Stability | Better than ± 0.2 % FS/± 0.2 % FS |
Measurement Accuracy | Better than ± 0.1 % FS @ 50 to 85°F Better than ± 0.2 % FS @ full temperature range |
Long Time Stability | Better than ± 0.1 % FS per year |
Materials | Housing: stainless steel (AISI 316 L) Diaphragm: 96,6 % Al2O3/Cell Packing: Viton |
Supply Voltage | 10 to 30 V DC |
Output Signal | 2-wire 4 - 20 mA (passive transmitter) |
Cable | 2 × AWG 20 (pressure) + 5 x AWG 25 (data), shielded, PUR insulation |
Cable Length | 39' |
Enclosure | NEMA 6P, withstands a static pressure equal to max. overpressure. |
Safety Approvals | UL Class I, II and II, Division 1 Hazardous (Classified) Locations, UL 913 ATEX: II 2G EEx ia IIC T4...T6 |
Test Approvals | CE directive: 89/336/EEC, 2004/108/EC, 1999/EC EN 61326-1, 61326-2-3, EN 55011, 61000-4-2, 61000-4-3, 61000-4-4, 61000-4-5,61000-4-6, 61000-4-8 EN 61000-6-4 2007-02-19, EN 61000-6-2 2005-09-08 |